Electrical Characteristics (continued). PARAMETER. SYMBOL. CONDITIONS. MIN. TYP. MAX. UNITS. IR LED CHARACTERISTICS (Note 3)

Electrical Characteristics (continued) PARAMETER SYMBOL CONDITIONS MIN TYP MAX UNITS IR LED CHARACTERISTICS (Note 3)

Modelllinie für dieses Datenblatt

Textversion des Dokuments

MAX30101 High-Sensitivity Pulse Oximeter and Heart-Rate Sensor for Wearable Health
Electrical Characteristics (continued)
(VDD = 1.8V, VLED+ = 5.0V, TA = +25°C, min/max are from TA = -40°C to +85°C, unless otherwise noted. Typical values are at TA = 25°C.) (Note 1)
PARAMETER SYMBOL CONDITIONS MIN TYP MAX UNITS IR LED CHARACTERISTICS (Note 3)
LED Peak Wavelength λP ILED = 20mA, TA = +25°C 870 880 900 nm Full Width at Half Max Δλ ILED = 20mA, TA = +25°C 30 nm Forward Voltage VF ILED = 20mA, TA = +25°C 1.4 V Radiant Power PO ILED = 20mA, TA = +25°C 6.5 mW
RED LED CHARACTERISTICS (Note 3)
LED Peak Wavelength λP ILED = 20mA, TA = +25°C 650 660 670 nm Full Width at Half Max Δλ ILED = 20mA, TA = +25°C 20 nm Forward Voltage VF ILED = 20mA, TA = +25°C 2.1 V Radiant Power PO ILED = 20mA, TA = +25°C 9.8 mW
GREEN LED CHARACTERISTICS (Note 3)
LED Peak Wavelength λP ILED = 50mA, TA = +25°C 530 537 545 nm Full Width at Half Max Δλ ILED = 50mA, TA = +25°C 35 nm Forward Voltage VF ILED = 50mA, TA = +25°C 3.3 V Radiant Power PO ILED = 50mA, TA = +25°C 17.2 mW
PHOTODETECTOR CHARACTERISTICS (Note 3)
Spectral Range of Sensitivity Λ > 30% QE QE: Quantum Efficiency 640 980 nm Radiant Sensitive Area A 1.36 mm2 Dimensions of Radiant Sensitive Area L x W 1.38 x 0.98 mm x mm
INTERNAL DIE TEMPERATURE SENSOR
Temperature ADC Acquisition Time TT TA = +25°C 29 ms Temperature Sensor Accuracy TA TA = +25°C ±1 °C Temperature Sensor Minimum Range TMIN -40 °C Temperature Sensor Maximum Range TMAX 85 °C
DIGITAL INPUTS (SCL, SDA)
Input Logic-Low Voltage VIL 0.3 x VDD V Input Logic-High Voltage VIH 0.7 x VDD V Input Hysteresis VHYS 0.5 x VDD V Input Leakage Current IIN ±0.1 ±1 µA Input Capacitance CIN 10 pF
DIGITAL OUTPUTS (SDA, INT)
Output Low Voltage VOL ISINK = 3mA 0.4 V www.maximintegrated.com Maxim Integrated │ 4